Instruction Manual
Wiring Rosemount 396RVP
LIQ-MAN-396R-396RVP
March 2017
Wiring Rosemount 396RVP
27
3.1
Wiring for Rosemount 396RVP
Make electrical connections as shown on Figures 4-1 through 4-20 using the following guidelines.
For wiring Rosemount 396R, see Section 3.
1.
Pay particular attention to the analyzer or transmitter model number when following
details on the wiring diagrams to ensure that the connections are made to the proper
terminals.
2.
The Rosemount 396RVP uses a mating VP cable. See attached wiring sheet for wire
functions of the cables and wiring diagrams to various transmitters.
3.
The maximum distance from the sensor to the transmitter is 15 ft without an integral
preamplifier.
4.
Signal cable should be run in a dedicated conduit and should be kept away from AC power
lines.
Section 3: Wiring Rosemount 396RVP
For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of
the sensor should be connected to the outer braided shield of the extension cable. The outer braid of the
extension cable to the instrument must be terminated at earth ground or by using an appropriate metal
cable gland fitting, that provides a secure connection to the instrument cable.