The following procedure outlines the major steps needed to configure a
Tri-Loop. For more detailed information, use the Configurator on-line
help screens.
1. Double click on the HART Tri-Loop Configurator Software icon.
2. Select Tri-Loop, Connect to Tri-Loop to display the connect
screen, then connect to a Tri-Loop.
NOTE
If you are unable to connect to a Tri-Loop, see Table 4-1 on page 4-1 for
suggestions.
3. Select Configuration, Configure Channels to display the HART
Tri-Loop Configuration Screen (Figure 3-3).
a. Click Channel 1, and select the process variable for this
channel, desired range, and
exact
Model 3095 MV units.
If required, click on the Enabled box so that the Tri-Loop
analog output will be enabled.
b. (Optional) Repeat step a for Channels 2 and 3.
c. Click device information, enter the desired tag, descriptor, and
message information, then select OK.
4. Select Configuration, Save to Tri-Loop to send the configuration
to a Tri-Loop.
5. Select Configuration, Open from Tri-loop to verify that the
configuration was correctly received. The Tri-Loop configuration
information will be displayed in the main window of the HART
Tri-Loop Configurator software.
6. If necessary, perform an analog output trim (Tri-Loop, Trim
Outputs) for each enabled Tri-Loop analog output.
7. Complete the connection between the Model 3095 MV analog
output and the Tri-Loop Burst Input as illustrated in Figure 3-4.
8. Use a multimeter or the control room equipment to verify that
each new Tri-Loop channel is transmitting the Model 3095 MV
process variables.
FIGURE 3-3. HART Tri-Loop
Configuration Screen.
30
95
-30
9
5
070
2
Summary of Contents for 333 HART Tri-Loop
Page 1: ...00809 0100 4754 English Rev BB Model 333 HART Tri Loop HART to Analog Signal Converter ...
Page 2: ......
Page 4: ......
Page 8: ...HART Tri Loop HART to Analog Signal Converter 1 2 ...
Page 26: ...HART Tri Loop HART to Analog Signal Converter 4 2 ...
Page 34: ...HART Tri Loop HART to Analog Signal Converter A 4 ...
Page 47: ......