Installing the HART Modem
After installing the 3095 Engineering Assistant, the HART modem configuration window
automatically appears when the 3095 Engineering Assistant is opened. If this window is
canceled, the HART modem may be installed manually with the following procedure.
1. Close AMS. From the Start Menu select
Programs
>
3095 Engineering Assistant
>
AMS Network
.
2. Select
Add
from the
AMS Network Configuration
window. A “
Select AMS Network
Component Type
” window will open.
3. Select
Hart Modem
.
4. Select
Install
.
5. Step through the Wizard. A prompt will inquire a COMM PORT. A typical default is
“COMM 1”.
6. Select
OK
. Close out of the
AMS Network Configuration
window.
7. Re-open the
HART Modem Properties
. Click on the
Connection
tab and set the Retry
Count to 5. Click the
Advanced
tab and check the
Multidrop
checkbox and set the
Address Range as 0 to 2.
8. The configuration change will activate when 3095 Engineering Assistant is launched.
NOTE
Exit from the program if a Palm Pilot HotSync or any other program is shared with the
COMM PORT.
3095 Engineering Assistant (EA) for F
OUNDATION
Fieldbus
1. Install the 3095 Engineering Assistant for F
OUNDATION
Fieldbus Program and
the F
OUNDATION
fieldbus communication card drivers.
a.Place CD-ROM disk 2 into the drive.
b.Browse and select 3095 Engineering Assistant for F
OUNDATION
Fieldbus software
application from Windows NT, 2000, or XP
c. Open the ReadMe.txt file and follow the instructions provided.
2. Install the F
OUNDATION
Fieldbus communication card.
a.Insert either the PCMCIA card or PCI card into the computer, following the
instructions provided with the communications card.
3. Connect the 3095 transmitter to computer and establish communications.
a.Connect the 9-pin communications cable into the F
OUNDATION
Fieldbus card port
located in the computer.
b.Connect communication wiring to connection ports labeled “D
+
” and “D
-
”.
c. Open the transmitter cover on the side marked “Field Terminals” and connect the
communication wires to the two 3095 transmitter terminals labeled “Fieldbus
Wiring”.
4. Select the 3095 Engineering Assistant for F
OUNDATION
Fieldbus from the program menu
or use the 3095 EA F
OUNDATION
fieldbus shortcut icon.
Summary of Contents for 3095MF Series
Page 29: ...Quick Installation Guide 00825 0100 4716 Rev DA October 2009 Rosemount 3095 MultiVariable 29 ...
Page 30: ...Quick Installation Guide 00825 0100 4716 Rev DA October 2009 Rosemount 3095 MultiVariable 30 ...
Page 31: ...Quick Installation Guide 00825 0100 4716 Rev DA October 2009 Rosemount 3095 MultiVariable 31 ...
Page 32: ...Quick Installation Guide 00825 0100 4716 Rev DA October 2009 Rosemount 3095 MultiVariable 32 ...