![Rosemount Analytical 54C Instruction Manual Download Page 23](http://html1.mh-extra.com/html/rosemount-analytical/54c/54c_instruction-manual_1487466023.webp)
3-9
MODEL 54 C
SECTION 3
PROGRAMING THE ANALYZER
This section explains configuration of the alarms. The designations
“A1”, “A2”, “A3” and “A4” refer to relays 1,2,3 and 4, respectively.
1.
From the program menu select configure menu and Enter (F4). With
the cursor on “Alarms”, press Enter (F4).
NOTE
After completing this section be sure to check the alarm set
points with function configuration (see
sary.
2.
*Delays relay deactivation after reading is no longer in the alarm 1 condition
Pressing More (F3) allows configuration of alarm 2 options.
3.
*Delays relay deactivation after reading is no longer in the alarm 2
condition
Pressing More (F3) allows configuration of alarm 3 options.
4.
*Pressing More (F3) proceeds to screen 5 if relay 3 is a process
alarm or screen 6 if relay 3 is a timer.
**Note: If A4 configuration is “Fault” the relay will activate in a fault
condition. To activate the complete diagnostics refer to
3.5 CONFIGURATION OF ALARM PARAMETERS
Display
Outputs
Alarms
Exit
More
Enter
A1:
Low
A1 Hys:
000 %
A1 Delay:
0 sec
Exit
More
Enter
A2:
High
A2 Hys:
000 %
A2 Delay:
0 sec
Exit
More
Enter
Menu Item
Options
A1 alarm 1: Logic
Low alarm/High alarm/
Off (not used)
A1 alarm 1: Hysteresis
0-25 % of full scale alarm
(deadband)*
alarm setpoint
A1 alarm 1: Delay time before relay activation
and deactivation (can prevent chatter)
0-99 sec
Menu Item
Options
A2 alarm 2: Logic
Low alarm/High alarm/
Off (not used)
A2 alarm 2: Hysteresis (Deadband)*
0-25 %
A2 alarm 2: Delay time before relay activation and
deactivation (can prevent chatter)
0-99 sec
Menu Item
Options
A3 alarm 3: Type
Timer/Process alarm*
A4 alarm 4 Logic
Fault/Off**
Relays 1-3 action in a hold or fault condition
Relays open/closed/none
A3:
Process
A4:
Fault
Relay default: Open