3-2
MODEL 54 C
SECTION 3
PROGRAMING THE ANALYZER
ITEM
RANGES
FACTORY SETTINGS
USER SETTINGS
B. Outputs (
Continued
3. Output 2
(a) Measurement Selection
Temperature/Conductivity
Temperature
_______
(b) Current Range
4-20 mA/0-20 mA
4-20 mA
_______
(c) Dampening
0-299 Sec
0 Sec
_______
(d) Hold Mode Value
Last value/Fixed value
Last value
_______
(value if (d) Fixed)
0-22 mA
21 mA
_______
(e) Fault value
0-22 mA
22 mA
_______
C. Alarms (
1. Alarm 1 (Process Alarm)
(a) Configuration
Low alarm/High alarm/Off
Low alarm
_______
(b) Hysteresis (deadband)
0-25% of setpoint
0%
_______
(c) Delay Time
0-99 sec
0 sec
_______
2. Alarm 2 (Process Alarm)
(a) Configuration
Low alarm/High alarm/Off
High alarm
_______
(b) Hysteresis (deadband)
0-25% of setpoint
0%
_______
(c) Delay Time
0-99 sec
0 sec
_______
3. Relay Action in Fault Mode (Alarms 1 - 3)
Open/Closed/None
Open
_______
4. Alarm 3
(a) Type
Process alarm/Timer
Process alarm
_______
(b) Settings if (a) is Process alarm
• Configuration
High alarm/Low alarm/off
High alarm
_______
• Hysteresis (deadband)
0-25% of setpoint
0%
_______
• Delay Time
0-99 sec
0 sec
_______
(c) Settings if (a) is Timer
• Interval between timer cycles
0-999.9 hr
24 hr
_______
• Number of relay activations per cycle
1-60
1
_______
• Relay on time per activation
0-999 sec
120 sec
_______
• Relay off time between activations
0-999 sec
1 sec
_______
• Recovery time after last activation
0-999 sec
600 sec
_______
5. Alarm 4 (configuration)
(a) Type
Fault/Off
Fault
_______
D. Security (
1. Level 1 (Lock all)
000-999
000 (no security)
_______
2. Level 2 (Lock all except Calibrate)
000-999
000 (no security)
_______
3. Level 3 (Lock all except Calibrate,
Alarm Setpoints and Rerange Outputs)
000-999
000 (no security)
_______
RTD Type
Pt100/Pt1000
Pt100
_______
Table 3-1. Conductivity Settings List
(Continued)