MANUFACTURER’S DECLARATION – ELECTROMAGNETIC
The Romed Infrared Forehead Thermometer is intended for use in the electromagnetic environment
specified below. The customer or the user of The Romed Infrared Forehead Thermometer should assure
that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance level
Electromagnetic environment
-guidance
Conducted
RF
IEC 61000-
4-6
Radiated
RF
IEC 61000-
4-3
3 Vrms 150kHz to
80MHz
6 Vrms in
ISM bands
3V/m 80MHz to
2.5GHz
385MHz-5786MHz
Test specification
for ENCLOSURE
PORT IMMUNITY
to RF wireless
communication
equipment(refer
to table 9 IEC
60601-1-2:2
014)
Not applicable
Not applicable
3V/m
80MHz to 2.5GHz
385MHz-5786MHz
Test specification
for ENCLOSURE
PORT IMMUNITY
to RF
wireless
communication
equipment(refer to
table 9 IEC
60601-1-2:2014)
Portable and mobile RF communications
equipment should be used no closer to any
part of The Romed Infrared Forehead
Thermometer , including cables, than the
recommended
separation distance calculated from the
equation application to the frequency of the
transmitter.
Recommended separation distance
Where p is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
metres(m).
Field strengths form fixed RF transmitters, as
determined by an electromagnetic site survey,
a should be less than the
compliance level in each frequency range. b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and
reflection from structures, objects and people.
a Field strengths from transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which The
Romed Infrared Forehead Thermometer is used exceeds the applicable RF compliance level above, The
Romed Infrared Forehead Thermometer should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating The
Romed Infrared Forehead Thermometer .
b Over the frequency range 150kHz to 80 MHz, filed strengths should be less than 3V/m.