77-014 - Yasui Rotating Burnout Furnace 37+
63
E0800
[ Door open ] Warning
- Close door.
- Door limit switch defective.
Check DOOR 2 LED shows reaction or not
at maintenance window.
E0801
[ Door open ] Error
- Close door and push START switch.
- Door limit switch defective. Check DOOR 2 LED shows reaction or not at maintenance window.
E0900
[ Blackout ]
- Informs blackout record of longer than 5
minutes.
E1300
[ Turn table does not rotate ]
- Check motor rotating at maintenance window.
- Check photo sensor 2 ON/OFF when switch
is ON or OFF.
E1605
[ Thermocouple disconnection of temperature
rise safety device ]
- Check if LED on SSR activated or not by
turning ON at maintenance window.
- Replace thermocouple.