Software

R&S

®

TS-PSM3

31

User Manual 1178.2780.02 ─ 01

Figure 7-1: Softpanel R&S TS-PSM3

7.3 Programming with GTSL libraries

/*
This example connects an external power supply to a load via TS-PSM3
channels 9 and 10. The power supply is connected to CH9_COM and
CH10_COM at the rear of the module TS-PSM3 and the load is connected
to CH9_NO and CH10_NO at the front of the module. The sense-lines of
the power supply are routed to the load via TS-PSM3 channels 1 and 2.
The current through channel 10 is determined indirectly by measuring
the voltage at the current-voltage-converter of channel 10.
Later on, the load is set to standby mode and the low standby current
is measured directly via the ampere meter of TS-PSAM. Therefore
the power relay of channel 9 is opened and the standby current is
bypassed via matrix relays to the analog bus and TS-PSAM module.

Error handling is not considered in this sample in order to keep it

 

Programming with GTSL libraries

Summary of Contents for TS-PRIO3

Page 1: ...R S TS PSM3 High Power Switch Module User Manual User Manual 1178 2780 02 01 K 2 Test Measurement...

Page 2: ...9 41 29 12 164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trad...

Page 3: ...e of the product The product is used for its designated purpose if it is used in accordance with its product documentation and within its performance limits see data sheet documentation the following...

Page 4: ...ation item 7 Signal words and their meaning The following signal words are used in the product documentation in order to warn the reader about risks and dangers Indicates a hazardous situation which i...

Page 5: ...afety is not observed either at all or to the extent necessary electric shock fire and or serious personal injury or death may occur 1 Prior to switching on the product always ensure that the nominal...

Page 6: ...t be fuse protected in such a way that anyone who has access to the product as well as the product itself is adequately protected from injury or damage 14 Use suitable overvoltage protection to ensure...

Page 7: ...masks and protective clothing must be worn 7 Laser products are given warning labels that are standardized according to their laser class Lasers can cause biological harm due to the properties of the...

Page 8: ...n permitted 5 If a cell develops a leak the fluid must not be allowed to come into contact with the skin or eyes If contact occurs wash the affected area with plenty of water and seek medical aid 6 Im...

Page 9: ...s substances or fuels can cause health problems and lead to environmental damage For additional information about environmental protection visit the Rohde Schwarz website Instrucciones de seguridad el...

Page 10: ...el uso del producto Deber n tenerse en cuenta todas las dem s informaciones de seguridad como p ej las referentes a la protecci n de personas que encontrar n en el cap tulo correspondiente de la docu...

Page 11: ...l producto se emplea de forma sin nima el t rmino CUIDADO Las palabras de se al corresponden a la definici n habitual para aplicaciones civiles en el rea econ mica europea Pueden existir definiciones...

Page 12: ...la tensi n preseleccionada en el producto coincida con la de la red de alimentaci n el ctrica Si es necesario modificar el ajuste de tensi n tambi n se deber n cambiar en caso dado los fusibles corre...

Page 13: ...efectuadas por un electricista especializado 13 En el caso de dispositivos fijos que no est n provistos de fusibles interruptor autom tico ni otros mecanismos de seguridad similares el circuito de ali...

Page 14: ...presario operador tiene la obligaci n de evaluar y se alizar las reas de trabajo en las que exista un riesgo elevado de exposici n a radiaciones 6 Tenga en cuenta que en caso de incendio pueden despre...

Page 15: ...r ni fuego Debe evitarse el almacenamiento a la luz directa del sol Las celdas y bater as deben mantenerse limpias y secas Limpiar las conexiones sucias con un pa o seco y limpio 3 Las celdas o bater...

Page 16: ...vos el ctricos usados no se deben desechar con los residuos dom sticos sin clasificar sino que deben ser recogidos por separado Rohde Schwarz GmbH Co KG ha elaborado un concepto de eliminaci n de resi...

Page 17: ...ed for exam ple according to the ISO9001 and ISO14001 management systems Environmental commitment Energy efficient products Continuous improvement in environmental sustainability ISO14001 certified en...

Page 18: ...ument up to date and to be informed about new application notes related to your instrument please send an e mail to the Customer Support Center stating your instrument and your wish We will take care...

Page 19: ...s via apertures on the front plate of the module to the terminals of the channels A soft panel is available for operating the R S TS PSM3 Control is via an IVI switch driver 1 2 Safety advice To preve...

Page 20: ...of R S TS PRIO3 rear I O module possible Feed through of all low power channels and high power channels 9 to 14 to the rear of the R S TS PCA3 chassis by means of optional R S TS PK04 cable set possi...

Page 21: ...igure 2 1 shows the view of the High power Switch Module R S TS PSM3B variant 02 Figure 2 1 View of the R S TS PSM3B variant 02 Figure 2 2 shows the view of the High power Switch Module R S TS PSM3 va...

Page 22: ...le R S TS PSM3 I V I V 4 x 30 A Design CH9 to CH16 4 x 2 A Design CH1 to CH8 X20 Row B Row A X1 X2 X30 CHx_COM CHx_COM CHy_COM CHy_COM CHx_NO CHy_NO X10 Controller CAN 5V Processor Relay Driver RAM An...

Page 23: ...board to which all relays and terminal blocks are attached a digital circuit board mounted on it on which the processor and the FRAM are located and a plate mounted above it which stabilises the modu...

Page 24: ...Construction R S TS PSM3 8 User Manual 1178 2780 02 01 Figure 4 1 R S TS PSM3 variant 03 Front with VPC R S TS PSM3...

Page 25: ...n nector is located on the upper edge of the base board in the middle of the module Figure 4 2 R S TS PSM3 Rear to backplane of the chassis In contrast to the R S TS PSM3B variant 02 the cabling of th...

Page 26: ...Connectors Table 4 1 Connectors on the R S TS PSM3 Abbreviation Use X1 Interface of channels 9 to 14 with optional R S TS PRIO3 rear I O module for use in the R S TS PWA3 or for optional R S TS PK04 c...

Page 27: ...he R S TS PSM3 is accessed via the interface PWR green Power Lights up when all supply voltages are present LED Test When voltage is powered up all three LED s light up for around 1 second This ensure...

Page 28: ...S TS PRIO3 backplane side to the R S TS PSM3 The X12 and X14 connectors are located on the side accessible to the user rear of the R S TS PWA3 Here loads or supply units can be connected to the low p...

Page 29: ...connectors can take wires with a maximum of 0 5 mm2 crosssection surface while the X14 connector is suitable for wires with a maximum of 6 mm2 cross section surface The metal plate attached to the fro...

Page 30: ...the load or voltage source X14 Interface of high power channels 9 to 16 with the load or voltage source X20 Interface of channels 1 to 8 with the R S TS PSM3 4 3 R S TS PK04 If the optional R S TS PK0...

Page 31: ...TSVP R S TS PCA3 4 3 1 Mechanical construction In Figure 4 9 the installation of the R S TS PK04 cable set in the CompactTSVP R S TS PCA3 cTSVP and its connections to the R S TS PSM3 here shown in sim...

Page 32: ...SM3 X3 und X4 Interface of the high power channels 9 to 14 and low power channels 1 to 8 with the load or voltage source X20 Interface of low power channels 1 to 8 with the R S TS PSM3 So that the max...

Page 33: ...characteristics are available for each high current channel of the R S TS PSM3 These values are used for calculating the actual current through the channel The amount of current through a channel can...

Page 34: ...example shown below Figure 5 1 the voltage supply is connected to the COM lines on the rear of the R S TS PSM3 This can be done via the R S TS PRIO3 optionally installed in the R S TS PWA3 or the R S...

Page 35: ...agram above the lines with sense voltage are blue and the paths over which the current voltage converter s voltage are mea sured are coloured green 5 1 3 2 Current measurement with load in standby mod...

Page 36: ...Functional description R S TS PSM3 20 User Manual 1178 2780 02 01 Figure 5 2 Application example current measurement with load in standby mode R S TS PSM3...

Page 37: ...ponents on the underneath of the module are not damaged The plate is unscrewed via three screws on the top of the module and two screws on the front plate of the module The digital circuit board is fi...

Page 38: ...2 User Manual 1178 2780 02 01 Figure 6 1 Position of the connecting terminals on the base board 6 1 2 Mounting the terminal wires high power channels see Figure 6 2 to Figure 6 4 Channel wiring of the...

Page 39: ...y a light lever action and the screwdriver pushed into the terminal as far as it will go Figure 6 2 Opening the terminal contact CH10_COM Figure 6 3 The wire can be mounted in the lower round opening...

Page 40: ...745 PCB terminal 5 mm 0 08 4 mm2 6 1 3 Mounting the terminal wires low power channels see Figure 6 5 For the terminals of low power channels 1 to 8 the plastic fish plate on the upper side of the term...

Page 41: ...e finely stranded More information on this terminal can be obtained from the manufacturer WAGO The terminal is labelled as follows Series 2060 SMD PCB terminal with activation trigger 4 mm 0 34 0 75 m...

Page 42: ...K04 cable set even to the rear of the R S TS PCA3 CompactTSVP to con nectors X3 and X4 To install plug in module High Power Switch Module R S TS PSM3 proceed as fol lows Shut down and switch off the R...

Page 43: ...n the HF seal of the power supply unit which may be warped and the R S TS PSM3 For safety reasons it is recommended that slots 1 2 in the R S TS PWA3 PowerTSVP may not be used for the R S TS PSM3 To i...

Page 44: ...te Rear I O slot corresponding to R S TS PSM3 Loosen the screws and remove the appropriate front plate section from the Pow erTSVP chassis Push in the R S TS PRIO3 using moderate force The R S TS PRIO...

Page 45: ...t Platforms R S CompactTSVP PCA3 and R S PowerTSVP PWA3 are designed for a maximum voltage of 125 V The R S TS PSM3 module is suitable for voltages of up to 30 V DC and must only be used accordingly F...

Page 46: ...es are installed Table 7 1 Driver Installation R S TS PSM3 Module Path Comment rspsm3 dll GTSL directory Bin Driver rspsm3 hlp rspsm3 chm GTSL directory Bin Help files rspsm3 fp GTSL directory Bin Lab...

Page 47: ...sense lines of the power supply are routed to the load via TS PSM3 channels 1 and 2 The current through channel 10 is determined indirectly by measuring the voltage at the current voltage converter o...

Page 48: ...ll SFTPrefix SFTMPSM3 device PSAM Description TS PSAM Module in Frame 1 Slot 6 Type PSAM ResourceDesc PXI3 12 INSTR Frame 1 Slot 6 DriverDll rspsam dll DriverPrefix rspsam DriverOption Simulate 0 Rang...

Page 49: ...CH2_COM PSM3 CH2_COM PSM3_CH9_NO PSM3 CH9_NO PSM3_CH9_COM PSM3 CH9_COM PSM3_CH10_NO PSM3 CH10_NO PSM3_CH10_COM PSM3 CH10_COM PSM3_CH10_IV PSM3 CH10_IV PSM3_LABD1 PSM3 LABD1 include ansi_c h include us...

Page 50: ...nd DMM_LO to local analog bus lines b1 and d1 ROUTE_Execute 0 residRoute DMM_HI LABb1 DMM_LO LABd1 errorOccurred errorCode errorMessage connect the current voltage converter of TS PSM3 channel 10 to a...

Page 51: ...m 1 0 1 voltageResult resultsCount errorOccurred errorCode errorMessage get the current value which corresponds to the measured voltage value at TS PSM3 channel 10 rspsm3_GetCalculatedCurrent psm3Sess...

Page 52: ...on TS PSM3 channel 9 wait for debounce for all switch modules ROUTE_Execute 0 residRoute PSM3_CH9_NO PSM3_CH9_COM errorOccurred errorCode errorMessage measure the current throught the ampere meter of...

Page 53: ...on the module LED faulty No LED s light up No 5 V supply 8 2 Startup test The startup test runs in parallel with the LED test The following information can be drawn from the different LED display sta...

Page 54: ...f the modules in the system is ensured by measurements via the analogue measurement bus For information on starting the self test and the order of required work steps please consult the GTSL Software...

Page 55: ...ia the connector the two contacts of a high power channel must always be connected in parallel A contact is only designed for a maximum of 16 A Figure 9 1 R S TS PSM3 X1 connector view plug side Table...

Page 56: ...aximum current can be run via the connector the two contacts of a high power channel must always be connected in parallel A contact is only designed for a maximum of 16 A Figure 9 2 R S TS PSM3 X2 con...

Page 57: ...3 variant 03 X10 connector Table 9 3 R S TS PSM3 variant 03 X10 connector assignment Pin A B Pin A B 1 CH16_NO CH2_COM 9 CH16_COM CH6_COM 2 CH15_NO CH1_COM 10 CH13_COM CH5_COM 3 CH13_NO CH4_COM 11 CH1...

Page 58: ...ion R S TS PSM3 42 User Manual 1178 2780 02 01 Pin A B Pin A B 7 CH9_NO CH4_NO 15 CH15_COM CH8_NO 8 CH10_NO CH3_NO 16 CH10_COM CH7_NO 9 1 4 X20 connector Figure 9 4 R S TS PSM3 X20 connector view plug...

Page 59: ...GA2 GA3 GA4 21 GA5 20 5V GND 5V 19 5V GND 18 CAN_EN 17 GND 16 GND 15 5V GND 14 13 12 CH4_COM CH2_COM 11 10 CH3_COM CH1_COM 9 8 CH6_COM CH8_COM 7 6 CH5_COM CH7_COM 5 4 3 RSA0 RRST GND RSDO 2 RSDI RSA1...

Page 60: ...PSM3 44 User Manual 1178 2780 02 01 9 1 5 X30 connector Figure 9 6 R S TS PSM3 X30 connector view plug side Pin E D C B A 7 6 GND 5 ABc1 ABa1 4 ABb1 3 ABc2 2 ABa2 1 ABd2 Figure 9 7 R S TS PSM3 X30 con...

Page 61: ...1 X1 connector Figure 9 8 R S TS PRIO3 X1 connector view plug side Table 9 4 R S TS PRIO3 X1 connector assignment Pin Channel Pin Channel 7 CH13_COM 1 CH13_COM 8 CH11_COM 2 CH11_COM 9 CH9_COM 3 CH9_C...

Page 62: ...er Manual 1178 2780 02 01 9 2 2 X2 connector Figure 9 9 R S TS PRIO3 X2 connector view plug side Table 9 5 R S TS PRIO3 X2 connector assignment Pin Kanal Pin Kanal 7 CH16_COM 1 CH16_COM 8 2 9 3 10 4 1...

Page 63: ...3 X20 connector Figure 9 10 R S TS PRIO3 X20 connector view plug side Pin A B C D E Table 9 6 R S TS PRIO3 X20 connector assignment Pin A B C D E 22 21 20 19 18 17 16 15 14 13 12 CH2_COM CH4_COM 11 10...

Page 64: ...02 01 Pin A B C D E 5 4 3 2 1 9 2 4 X12 connector Figure 9 11 R S TS PRIO3 X12 connector view plug side Table 9 7 R S TS PRIO3 X12 connector assignment Pin Channel 1 CH1_COM 2 CH2_COM 3 CH3_COM 4 CH4...

Page 65: ...1178 2780 02 01 9 2 5 X14 connector Figure 9 12 R S TS PRIO3 X14 connector view plug side Table 9 8 R S TS PRIO3 X14 connector assignment Pin Kanal 1 CH13_COM 2 CH11_COM 3 CH9_COM 4 CH14_COM 5 CH12_C...

Page 66: ...contacts of a high power channel must always be connected in parallel A contact is only designed for a maximum of 16 A Figure 9 13 X3 connector rear of R S TS PCA3 or R S TS PWA3 Table 9 9 X3 connecto...

Page 67: ...of a high power channel must always be connected in parallel A contact is only designed for a maximum of 16 A Figure 9 14 X4 connector rear of R S TS PCA3 or R S TS PWA3 Table 9 10 X4 connector assign...

Page 68: ...0 Specifications The technical data of the High Power Switch Module R S TS PSM3 are shown in the corresponding data sheets In the event of any discrepancies between date in this user manual and techni...