Application
R&S
®
TS-PSM3
4
User Manual 1178.2780.02 ─ 01
For further details on operation with hazardous voltages, see .
tions for operation with voltages dangerous to the touch"
1.3 Characteristics
Table 1-1: Characteristics R&S TS-PSM3
Characteristics R&S TS-PSM3
Deployable in the R&S TS-PCA3 (R&S Compact TSVP) and R&S TS-PWA3 (R&S Power TSVP) chassis
Control via CAN bus
Eight high power channels for currents up to 30 A
Eight low power channels for currents up to 2 A
Switching of voltages up to 30 V
Current measurement capabilities using shunt resistors or currentvoltage converters
Feed-through of all channels to the rear of the R&S TS-PWA3 chassis by means of R&S TS-PRIO3 rear
I/O module possible
Feed-through of all low power channels and high power channels 9 to 14 to the rear of the R&S TS-PCA3
chassis by means of optional R&S TS-PK04 cable set possible
Support via R&S Signal Routing Library TS-LSRL
Self-test capability
Soft panel for interactive operation
IVI switch driver available
In permanent operation of the 30 A relays with low switching voltages and low switch-
ing currents there is a risk of an increase in the contact resistance due to a forming film
on the contact surfaces. It is therefore recommended to regularly switch the contacts
under load (eg. 10 A / 12 V).
Table 1-2: R&S TS-PRIO3 characteristics
R&S TS-PRIO3 characteristics
Direct routing of all low-power channels to the rear panel of the basic unit R&S TS-PWA3.
Direct routing of all high-power channels to the rear panel of the basic unit R&S TS-PWA3.
Characteristics