Instrument Function
R&S
®
SMF100A
228
Operating Manual 1167.2319.02 ─ 12
Remote command:
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
Level Offset State- Power Analysis
Activates a level offset at the sensor input. Set the appropriate value in the entry field
on the right, see
Remote command:
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet:STATe
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet:STATe
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:OFFSet:STATe
Level Offset - Power Analysis
Sets the level offset at the sensor input. To consider the value, activate the offset with
Level Offset State- Power Analysis "
Remote command:
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:OFFSet
:SENSe<ch>[:POWer]:SWEep:POWer[:SENSor]:OFFSet
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:OFFSet
Use Separate Frequency- Power Analysis
This setting is offered for measurements with DUTs that change the measurement fre-
quency (like modulators), thus changing the input frequency of the sensor.
The dialog differs depending on the measurement modes:
●
Power versus frequency measurement
Activates the use of a different frequency range other than the set signal generator
frequency range for the measurement. The separate minimum and maximum fre-
quency values are set below.
The x-scale of the diagram can be adjusted to the separate frequency range with
functions "Use as X Scale" and "Map to X Scale".
●
Power versus power measurement / Power versus time measurement
Activates the use of a different frequency other than the set signal generator fre-
quency for the measurement. The separate frequency value is set in the entry win-
dow below.
RF Level