R&S NRP Power Sensors Tour
R&S
®
NRP Series
24
Getting Started 1419.0170.02 ─ 10
Power sensor
Male connector
Matching female con-
nector
Tightening torque
R&S NRP50TN
R&S NRP67T
1.85 mm
R&S NRP67TN
R&S NRP110T
1.0 mm
1.0 mm
0.23 Nm (2'' lbs)
Table 5-5: R&S
NRPxxTWG RF connector characteristics
Power sensor
Matching female
connector
Male connector
Frequency range
Tightening tor-
que
R&S NRP75TWG
WR-15
WR-15
50
GHz to 75
GHz
0.58 Nm (5'' lbs)
R&S NRP90TWG
WR-12
WR-12
60
GHz to 90
GHz
R&S
NRP110TWG
WR-10
WR-10
75
GHz to 110
GHz
*It is recommended to use the torque wrench of the R&S tool with the part number 1175.2014.02.
5.2
Trigger I/O Connector
The trigger I/O is a connector of SMB type.
It is used as an input for signals if the trigger source parameter is set to
EXTernal2
. It
is used as an output for trigger signals if the sensor is operated in the trigger master
mode.
Further information:
●
For details on configuration, see the user manual.
5.3
Host Interface
The host interface connector is used for establishing a connection between the power
sensors and a USB host or a supported Rohde
&
Schwarz instrument. For this pur-
pose, an external cable is needed. Two types of cables are available:
●
R&S NRP
‑
ZKU cable with a USB connector, for connecting the power sensor to a
USB host device (R&S order number 1419.0658.xx).
●
R&S NRP
‑
ZK6 cable with a push-pull type connector, for connecting the power
sensor to a base unit, R&S NRX or R&S NRP2, or other Rohde
&
Schwarz prod-
ucts with the round connector (R&S order number 1419.0664.xx).
These cables can be obtained in different lengths up to 5 meters.
Host Interface