Inspection and Maintenance
R&S
®
ATS1000
66
Instructions Handbook 1179.2983.02 ─ 01
The test ensures that the interlocks are functioning properly. See
Yearly safety check
This check is restricted to Rohde
&
Schwarz
.
Due to normal wear, the performance of any system can degrade over time. This per-
formance degradation can also impair system safety. To prevent any risks, we recom-
mend a regular safety and performance check of the chamber once a year.
8.3
Preparing the Chamber for Maintenance
Before carrying out any of the maintenance tasks described in
on page 66, perform the following steps.
1. Make sure that nobody uses the chamber during maintenance.
Take the steps stipulated at your company to prevent use of the chamber.
Chapter 7.3.3, "Opening the Door"
3. Deactivate the chamber as described in
Chapter 7.2, "Deactivating the Chamber"
on page 43.
Deactivation prevents any positioner movement that can lead to injuries during
maintenance activities.
4. If you need to move the chamber to a different location for maintenance, follow the
Chapter 5.1, "Moving the Chamber"
8.4
Performing Maintenance Tasks
The recommended intervals are listed in
8.4.1
Daily Functional Check
To check the door's gasket
1. Check the door's gasket for soiling, damage and wear. For information on prolong-
ing the service life of the gasket, see
2. If the gasket is soiled, clean it as described in
Chapter 8.4.3.1, "Cleaning the Gas-
3. If the gasket has visible damage or wear, contact Rohde
&
Schwarz to have it
replaced; see
Chapter 9.3, "Contacting Customer Support"
Performing Maintenance Tasks