Machine Overview
R&S
®
ATS1000
24
Instructions Handbook 1179.2983.02 ─ 01
Door interlock
Figure 4-6: Upper interlock key for the door locking mechanism
1 = Male upper interlock key
2 = Female upper interlock keyhole
The upper interlock system prevents the door locking mechanism from assuming the
locked position while the door is still open, which can lead to equipment damage.
Positioner interlock
Figure 4-7: Lower interlock key for the positioners
1 = Male lower interlock key
2 = Female lower interlock keyhole
The lower interlock system stops the positioners from moving while the door is open,
which can lead to personal injury.
This "open-door" icon in the NCD controller shows the interlock state:
●
If the icon is visible, it indicates that the door is open and the interlock is stopping
the positioners from moving.
●
If the icon is not visible, the door is closed and the interlock does not stop the posi-
tioners from moving.
Interlock Systems