Instrument Tour
R&S
®
NGP800
18
Getting Started 5601.5603.02 ─ 05
1 = Display with touch screen
2 = Rotary knob and back key
3 = Output and channel keys
4 = Output terminals (see
)
5 = Chassis ground terminal (4mm socket)
6 = Standby button
7 = USB connector
8 = Menu control keys
Display (1)
The display is a color TFT touch screen. Depending on the instrument models, up
to four channels are shown on the display. The respective measurement settings
and menu settings are displayed in the individual channel display area.
Two information status bars, providing the overall device operating mode and
channel settings of the instrument are located respectively at the device level
(top-right hand corner of the display area) and channel level (on top of individual
channel display area) of the instrument.
For a detailed description on-screen layout, see section "Display Overview" in the
User Manual.
Rotary knob and back key (2)
The rotary knob and back key are used for menu navigation and value adjustment
in the instrument.
For a detailed description on navigation, see section "Rotary Knob and Back Key"
in the User Manual.
Output and channel keys (3)
Depending on the instrument models, up to four channels and one output key are
provided to select individual channel and enable/disable the output(s).
Output terminals (4)
Two-channel instruments: NGP802 and NGP822 are equipped with 8 terminals
for outputs and remote sense connections. Four-channel instruments: NGP804,
NGP814 and NGP824 are equipped with 16 terminals for outputs and remote
sense connections.
For 32 V models, each output is capable to source 200 W of power at 0 V to 32 V
and maximum current of 20 A.
Overview of Controls