Instrument control
R&S
®
HMP Series
45
Getting Started 1178.6791.02 ─ 03
5. Turn the [rotary knob] to increase or decrease the value.
Alternatively, use the [up] and [down] keys.
6. Press the [rotary knob] to confirm the value.
Values adjusted with the keypad must be applied with the [Enter] key. Other-
wise, the instrument disables the edit mode without the changes taking
effect.
7.4
Remote control
In addition to operating the R&S
HMP directly on the instrument, it is also possi-
ble to operate and control it from a remote PC.
Remote Control Interfaces
The R&S
HMP supports various interfaces for remote control:
●
Dual Ethernet / USB interface, option R&S
HO732
●
Dual USB / RS232 interface, option R&S
HO720
●
GPIB Interface IEEE-488, option R&S
HO740
To control the instrument remotely, you need one of the listed options installed in
your R&S
HMP. Drivers and information on their installation are provided for
www.rohde-schwarz.com/product/hmp
www.rohde-schwarz.com/product/hmp4000
.
See
Chapter 4.8, "Connecting to LAN"
on page 24 for an example on how to set
up a LAN connection for remote control.
Remote control