Remote Commands
R&S
®
FSW
728
User Manual 1173.9411.02 ─ 19
Useful commands for occupied bandwidth measurements described elsewhere
Configuring the channel:
●
●
●
Defining search limits:
●
●
●
Performing the measurement:
●
●
Retrieving results:
●
●
●
Remote commands exclusive to occupied bandwidth measurements
[SENSe:]POWer:BANDwidth|BWIDth
...............................................................................728
[SENSe:]POWer:BANDwidth|BWIDth
<Percentage>
This command selects the percentage of the total power that defines the occupied
bandwidth.
Parameters:
<Percentage>
Range:
10 PCT to 99.9 PCT
*RST:
99 PCT
Example:
POW:BWID 95PCT
Manual operation:
See
11.5.5.2
Programming Example: OBW Measurement
This programming example demonstrates the measurement example described in
chapter 5.5.5, "Measurement Example"
on page 193 in a remote environment.
//-------------Configuring the Measurement-----------------------
*RST
//Resets the instrument
FREQ:CENT 800MHz
//Sets the center frequency to 800 MHz.
DISP:TRAC:Y:RLEV -10dBm
//Sets the reference level to -10 dBm.