![Rohde & Schwarz FSW13 User Manual Download Page 699](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847699.webp)
Remote Commands
R&S
®
FSW
699
User Manual 1173.9411.02 ─ 19
[SENSe:]POWer:ACHannel:PRESet
<Measurement>
This command determines the ideal span, bandwidths and detector for the current
power measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
Parameters:
<Measurement>
ACPower | MCACpower
ACLR measurement
CPOWer
channel power measurement
OBANdwidth | OBWidth
Occupied bandwidth measurement
CN
Carrier to noise ratio
CN0
Carrier to noise ration referenced to a 1
Hz bandwidth
Manual operation:
See
"Optimized Settings (Adjust Settings)"
See
See
[SENSe:]POWer:ACHannel:PRESet:RLEVel
This command determines the ideal reference level for the current measurement.
This automatic routine makes sure that the that the signal power level does not over-
load the R&S
FSW or limit the dynamic range by too small a S/N ratio.
To determine the best reference level, the R&S
FSW aborts current measurements
and performs a series of test sweeps. After it has finished the test, it continues with the
actual measurement.
To get a valid result, you have to perform a complete sweep with synchronization to
the sweep end. This is only possible in single sweep mode.
Usage:
Event
[SENSe:]POWer:TRACe
<TraceNumber>
This command selects the trace channel power measurements are performed on.
For the measurement to work, the corresponding trace has to be active.
Parameters:
<TraceNumber>
Range:
1 to 6
*RST:
1
Example:
POW:TRAC 2
Assigns the measurement to trace 2.
Configuring and Performing Measurements