Remote Commands
R&S
®
FSW
935
User Manual 1173.9411.02 ─ 19
CALCulate:MARKer:X:SLIMits[:STATe]
<State>
This command turns marker search limits on and off.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:X:SLIM ON
Switches on search limitation.
Manual operation:
See
"Search Limits (Left / Right)"
See
"Deactivating All Search Limits"
See
CALCulate:MARKer:X:SLIMits:LEFT
<SearchLimit>
This command defines the left limit of the marker search range.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Parameters:
<SearchLimit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
left diagram border
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation:
See
"Search Limits (Left / Right)"
See
CALCulate:MARKer:X:SLIMits:RIGHT
<SearchLimit>
This command defines the right limit of the marker search range.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Parameters:
<Limit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
right diagram border
Analyzing Measurements (Basics)