Measurement Examples
FSMR
1166.3311.12-01 1.6
E-1
Fig.1-3
RF power measurement after Zeroing
2. Measure the RF power in Watt
Connect the power sensor to the signal generator RF output
PWR METER
DISPLAY
LOG LIN
Linear display – Change the Unit to Watt
Press the
PWR METER
hotkey.
The POWER METER menu is displayed.
In the POWER METER menu press the
DISPLAY LOG / LIN
softkey.
The RF power unit is changed from dB to Watt.
Fig.1-4
RF power measurement in UNIT Watt