![Rohde & Schwarz FSMR User Manual Download Page 5](http://html.mh-extra.com/html/rohde-and-schwarz/fsmr/fsmr_user-manual_1477835005.webp)
FSMR Measurement
Examples
1166.3311.12-01 1.3
E-1
Fig.1-1 Display after selecting the default setup in receiver mode
The main default parameters are listed in the following table:
Table 1-1
Default parameters after preset in measuring recveiver mode
Parameter Parameter
Name
Value
Receiver frequency
Frequency
100 MHz
Reference Level
Ref Level
Manual -20 dBm
RF attenuation
RF ATT
Auto
Preamplifier Preamp
Off
Demodulation Demod
FM
Detector Det.
+/-
peak/2
Measurement time
Meas Time
100 ms
Trigger Trigger
Free
run
Measurement Examples
All of the following examples assume the standard settings for the measuring receiver. These are set
with the
PRESET
key
The described measurement applications are:
•
Measurement of the RF power of a signal using the Power Meter mode.
•
Measurement of low level signals using the Tuned RF Level mode.
•
Measurement of AM modulated signals and modulation distortion with the Modulation Analyzer mode.
•
Measurement of audio signals with the Audio Analyzer mode.