Remote Control
R&S
®
ESRP
631
User Manual 1175.7445.02 ─ 08
Example:
CALC:DELT2:SGR:Y:MAX
Activates and positions delta marker 2 on the peak level of the y-
axis.
CALC:DELT2:SGR:Y:MAX:BEL
Positions delta marker 2 on the next higher level found in the
diagram area below the current marker position.
Usage:
Event
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum:NEXT
This command positions the specified delta marker on the next peak level of the verti-
cal axis. The search includes frames above and below the current marker position. It
does not change the horizontal position of the delta marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m>
1...16
Selects the marker.
Example:
CALC:DELT2:SGR:Y:MAX
Activates and positions delta marker 2 on the peak level of the y-
axis.
CALC:DELT2:SGR:Y:MAX:NEXT
Positions delta marker 2 on the next higher level.
Usage:
Event
CALCulate<n>:DELTamarker<m>:SGRam:Y:MAXimum[:PEAK]
This command positions the specified delta marker on the current peak level of the ver-
tical axis. The search is performed over all frames. It does not change the horizontal
position of the deltamarker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m>
1...16
Selects the marker.
Example:
CALC:DELT2:SGR:Y:MAX
Activates and positions delta marker 2 on the peak level of the y-
axis.
Usage:
Event
Remote Commands in Receiver Mode