Rocket

® 

Quick User Guide

Rocket

®

 

PSU

™ 

Portable Suction Unit

8.

  Place the PSU

 over the filter unit, so the filter  

will fit into the filter recess under the suction unit.

9.

  Place the PSU

 onto the chest drain unit. With 

slight downward pressure rotate the locking ring 

clockwise (when facing the unit) so that it locks 

into position.

Lock

Unlock

Summary of Contents for PSU Portable Suction Unit

Page 1: ...Rocket Quick User Guide Rocket PSU Portable Suction Unit...

Page 2: ...ime 3 Attaching the suction unit to the chest drain bottle 4 5 Operating instructions 6 8 Changing suction levels 9 Information screens 10 11 Switching the unit off 11 Disconnecting the suction unit f...

Page 3: ...ort is on the back of the device It is protected and is behind the door marked 4 Charge the device for approximately 3hrs until the battery display is full Changing the clock time Please check the tim...

Page 4: ...r attach tubing set to bottle by inserting straw and cap into aperture and twisting clockwise until clicked shut Take care not to touch the straw 3 Once underwater seal has been established place bott...

Page 5: ...the PSU over the filter unit so the filter will fit into the filter recess under the suction unit 9 Place the PSU onto the chest drain unit With slight downward pressure rotate the locking ring clockw...

Page 6: ...perators guide During set up you can confirm or cancel your decisions using the three buttons at the side of the screen 2 Verify if the PSU is connected to a new patient THIS INFORMATION IS IMPORTANT...

Page 7: ...ilter as per the instructions re fit the PSU to the chest drain bottle and switch on the unit again to restart the set up process If you press you will progress to the next screen Please verify that y...

Page 8: ...tisfactory The levels of suction applied increase in 0 5KPa 5cmH2O and 3 75mmHg increments There is a gravity drainage setting of 0 25KPa When you are happy with the set level of suction press 6 You w...

Page 9: ...gether for longer than 2 seconds This will return you to the suction level setting screen Repeat instruction number 5 to change the level of suction required and instruction 6 to confirm When the suct...

Page 10: ...of use Flow is measured from 10L min Screen 2 30 Minute Flow low The screen shows up to 30 minutes of use Flow is measured from 0 5L min Screen 3 Previous Minutes Flow This screen shows the previous 1...

Page 11: ...ked to check the filter and then switch off the alarm and screen manually The PSU will continue at the programme suction level Switching the unit off When this has been completed turn the device off b...

Page 12: ...6988 www rocketmedical com ZDOCK293 Rev 05 02 2021 For all your Rocket PSU Portable Suction Unit service enquiries please contact Service Information Ultrasound Technologies Ltd Tel 44 0 1291 425425...

Reviews: