CRAFT™ Oocyte Aspiration Pumps – Service Manual
23
Copyright© 2012-19
Rocket Medical plc. .
All rights reserved.
ZDOCK226
Rev:9
25/02/19
8.
FINAL TESTING:
1. Final testing should be
completed with a
validated and
independently calibrated
electronic vacuum gauge
or similar device.
2. Required operating scale:
10-500mmHg in 1mmHg
increments
3. Attach the vacuum test
gauge as indicated in the
image below to the
external vacuum port
4. Attach the footswitch as
for normal operation.
5. Power device on.
6. Occlude the tube to the
test gauge.
7. Activate the Standard
Vacuum footswitch
–
observe the green light is
illuminated.
8. Rotate the control knob
clockwise
to increase to
the desired value on the
gauge. (See table below)
9. To decrease the value:
rotate anticlockwise.
10. When the desired value is
reached, release the
footswitch and tubing set
occlusion.
WARNING
:
Ensure the vacuum
levels are within the
specified accuracy
tolerance before
returning the pump to
clinical service.
11. Activate the Standard
Vacuum footswitch.
12. Record the vacuum
registered.
13. Repeat from #6 for each
of the test values
indicated in the table
(right).
Specified accuracy: ±10% of test value
Test
Value
Indicated
vacuum
Test Gauge
50mmHg
100mmHg
200mmHg