Rocket
CRAFT™ Oocyte Aspiration Pumps – Service Manual
10
Copyright© 2012-19
Rocket Medical plc. .
All rights reserved.
ZDOCK226
Rev:8
25/02/19
WARNING:
ELECTRIC SHOCK
HAZARD.
The equipment is to be
used only with electrical
systems complying with
all IEC, CEC and NEC
requirements.
WARNING:
ELECTRIC SHOCK
HAZARD.
Do not immerse the
device.
WARNING:
Device can cause
explosion in the
presence of flammable
gases.
4.6. SUPPLY VOLTAGE SELECTION
The device operates at a voltage 110V 60Hz or 220-240 VAC @ 50Hz. 0.12A depending
on model.
Ensure that the correct power cord is connected.
4.7. ELECTROMAGNETIC COMPATIBILITY
Rocket Craft Suction Pumps comply with the electromagnetic compatibility (EMC) limits for
medical devices as specified by IEC 60601-1-2:2001. These limits are designed to provide
a reasonable degree of protection against harmful interference found in typical medical
installations.
Medical electrical equipment requires special precautions regarding EMC and the device
must be installed, positioned and operated according to the instructions contained in this
manual to ensure continued electromagnetic compatibility.
The device must be operated according to the instructions contained in this manual to
ensure continued electromagnetic compatibility.
4.8.PACKAGING
The packaging has been designed to allow secure transportation of the pump and its
accessories.
After unpacking, re-assemble and retain the packaging for transport for servicing when
required.
4.9. POSITIONING and PLACEMENT of the DEVICE
Rocket
CRAFT™ Oocyte Aspiration Pumps must be placed on a secure, level surface,
away from sources of heat, water splashes, mists or cooling vents. Do not expose to direct
sunlight.
Do not expose to direct sunlight.
Do not expose to flammable gases.
Operating temperature Range: +5°C and +35°C