CRAFT™ Oocyte Aspiration Pumps – Service Manual
IMPORTANT
4. Reseat the pump head
gasket
P/N: GO1-001
5. Replace the valve head
assembly onto the
gasket, ensure the holes
are correctly aligned and
the gasket with seated
correctly.
6. Tighten the valve head
screws in a diagonal
pattern to ensure the
pump head is correctly
seated without distortion.
7. Re-fit the pump silencer
to the head module.