CRAFT™ Oocyte Aspiration Pumps – Service Manual
21
Copyright© 2012-19
Rocket Medical plc. .
All rights reserved.
ZDOCK226
Rev:9
25/02/19
7.2
Replacing the gauge in DUO-VAC & Suction Pump models:
1. Remove
the
upper
casing. (See Section 6)
2. Lift the front panel out of
the retaining grooves on
the lower case
3. Loosen and remove the
securing nuts or screws
(ringed right).
4. Draw the gauge through
the front panel whilst
supporting the mounting
plate.
5. Detach the gauge from
the centre tube by
carefully pulling the
tubing from the rear
connection.
6. Replace the gauge,
reconnect the centre
tubing and insert through
the front panel.
7. Replace gauge mounting
plate and tighten securing
nuts or screws.
Rocket Craft DUO-VAC gauge assembly
Rocket Craft Suction Pump gauge assembly