159
8
8.3 List of Errors
w
WARNING
After the occurrence of an Error, your Accu-Chek Spirit Combo
insulin pump is in STOP mode and insulin delivery is inter-
rupted. In order to maintain insulin delivery, you must act
immediately according to the instructions given for each
Error.
Error E1: CARTRIDGE EMPTY
The insulin cartridge is empty. You must change it immediately.
j
Press
f
twice to snooze and to confirm the Error.
j
Disconnect or remove the infusion set from the infusion site.
j
Change the cartridge and prime the infusion set.
j
Reconnect the infusion set or insert a new one.
j
Put your pump in RUN mode.
See section
3.5
Changing the cartridge and infusion set.
Error E2: BATTERY EMPTY
The battery is empty. You must replace it immediately.
w
WARNING
After inserting a new battery, always check that the time and
date of your Accu-Chek Spirit Combo insulin pump are set
correctly. Incorrect programming of the time and /or date may
cause incorrect insulin delivery. If you, your doctor or
healthcare team store and analyse your therapy data
electronically and the time and date of the devices used are
not set identically, the resulting data might not be meaningful.
man_05957613001_E_en.indb 159
02.03.2015 10:44:48
Summary of Contents for ACCU-CHEK Spirit Combo
Page 8: ...man_05957613001_E_en indb 6 02 03 2015 10 43 53...
Page 28: ...24 man_05957613001_E_en indb 24 02 03 2015 10 43 55...
Page 64: ...60 man_05957613001_E_en indb 60 02 03 2015 10 44 15...
Page 136: ...132 man_05957613001_E_en indb 132 02 03 2015 10 44 45...
Page 146: ...142 man_05957613001_E_en indb 142 02 03 2015 10 44 47...
Page 226: ...222 My notes man_05957613001_E_en indb 222 02 03 2015 10 45 07...