162
In rare cases, the cartridge itself can be the cause of the
occlusion. If an Error E4: OCCLUSION occurs again, you
must change the cartridge. Follow the steps listed below:
j
Press
f
twice to turn off and to confirm the Error.
j
Disconnect or remove the infusion set from your infusion site.
j
Change the cartridge and prime the infusion set, as explained
in chapter
2
.
j
Reconnect or insert a new infusion set at your infusion site.
j
Put your pump in RUN mode.
If the Error E4: OCCLUSION occurs again, prime your pump
without a cartridge. Follow the steps listed below:
j
Press
f
twice to turn off and to confirm the Error.
j
Disconnect or remove the infusion set from your infusion site.
j
Remove the cartridge, adapter and infusion set from your pump.
j
Carry out the INFUSION SET PRIMING function (See section
2.8
Priming the infusion set
)
without
the cartridge, adapter,
and infusion set inserted.
j
Rewind the piston rod.
j
Insert a new cartridge with the adapter and infusion set
connected.
j
Prime the infusion set.
j
Reconnect or insert a new infusion set at your infusion site.
j
Put your pump in RUN mode.
If the Error E4: OCCLUSION continues to occur, use the
alternative therapy plan provided by your doctor or
healthcare team. Contact your local pump support (details
on inside back cover) for more assistance.
man_05957613001_E_en.indb 162
02.03.2015 10:44:48
Summary of Contents for ACCU-CHEK Spirit Combo
Page 8: ...man_05957613001_E_en indb 6 02 03 2015 10 43 53...
Page 28: ...24 man_05957613001_E_en indb 24 02 03 2015 10 43 55...
Page 64: ...60 man_05957613001_E_en indb 60 02 03 2015 10 44 15...
Page 136: ...132 man_05957613001_E_en indb 132 02 03 2015 10 44 45...
Page 146: ...142 man_05957613001_E_en indb 142 02 03 2015 10 44 47...
Page 226: ...222 My notes man_05957613001_E_en indb 222 02 03 2015 10 45 07...