188
Start-up graph over the stabilization period
The start-up graph shows changes in the flow rate over the
stabilization time.
flo
w ra
te (ml
/ h)
0.012
0.006
0.004
0.002
0.000
0
120
240
360
600
1080
1200
1440
480
720
840
960
1320
0.010
0.008
time (min)
real flow
target flow
0.01 ml is equivalent
to 1 U (U100 insulin)
Bolus
For U100 insulin the maximum deviation of a maximum bolus is
≤ ± 5 % and the maximum deviation of a minimum bolus is
≤ ± 30 %*.
* The measurements were made according to IEC 60601-2-24:1998 with a
Accu-Chek TenderLink infusion set 17 / 110 at room temperature.
man_05957613001_E_en.indb 188
02.03.2015 10:44:51