32 • Calibrating Instruments
NOTE:
The screen shown above applies to a 4-channel GX-2009. If
your GX-2009 has less than 4 channels, the inactive
channels will not appear in the above screen and in any
screens where channels are displayed. All screens in this
manual assume a 4-channel GX-2009.
7. There are two exceptions to the sequence described in step 6
above. See the GX-2009 User Setup Program Operator’s Manual
for a description of the Cal Limit Display and Cal Limit Check
instrument parameters.
•
When Cal Limit Display is set to On and Cal. Limit Check is
set to Can’t Use, if the GX-2009 is due for calibration the
instrument will show a calibration expired failure screen and the
buzzer and LED arrays will pulse for a few seconds before
connecting to the SM-2009U.
The failure screen will remain on the display while the unit is
connected to the SM-2009U.
NOTE:
You can still use the SM-2009 with an instrument for bump
testing and calibration if this failure indication occurs, but you
will not be able to perform any operations that require the
display until a calibration is performed, the unit is turned off,
and then the unit is re-connected to the SM-2009U. If you
perform a bump test or calibration in this case, the only
indication of pass or fail will be the BUMP or CAL. LED.
FAIL
C-LIMIT