64 • Chapter 5: User Mode and Calibration
GasWatch 3 for CO and O
2
Setting the Alarm Points
1. After entering the
ALARM-P
menu, press AIR to scroll to the instrument channel whose
alarm points you want to change..
2. Press and release POWER/MODE.
3. The Warning setpoint for the channel flashes. In the example below, the CO channel is
selected.
4. Use AIR to adjust the Warning setpoint. Alarm setpoint limitations are shown below.
5. Press and release POWER/MODE to save the setting.
6. Repeat Step 4 and Step 5 for the Alarm, Alarm H, STEL and TWA settings.
7. The instrument returns to the channel selection screen.
8. Repeat Step 1 - Step 7 to change the alarm points for the other channel.
9. See “ESCAPE” on page 65 to return to User Mode.
Channel
Alarm Point Limitations
CO
20 ppm
WARNING
ALARM
ALARM H
2,000 ppm
O
2
• 0.0%
ALARM
WARNING
20.0%
• 21.8%
ALARM H
40.0%
DEF . ALMP
ESCAPE
AIR
ALARM--P
CO
ppm
ALARM--P
O2
%
AIR AIR AIR
WARNING
25
CO
ppm
ALARM
50
CO
ppm
ALARM H
1200
CO
ppm
STEL
200
CO
ppm
TWA
25
CO
ppm
ALARM--P
CO
ppm
For all settings:
1) AIR to adjust
2) POWER/MODE
to save and
continue