16 • Chapter 3: Measuring Mode
GasWatch 3 for CO and O
2
4. If
BP.RMDR
is set to
ON
(factory setting is
OFF
) and a bump test is due, the screen that
appears next depends on how
BP.EXPD
is set in User Mode (see pg.62). The three
possible screens are described below. If a bump test is not due, the instrument displays the
number of days left until a bump test is due.
5. The Date/Time Screen appears for a few seconds.
BP.EXPD
set to
CONFIRM
(factory setting)
BP.EXPD
set to
CANT.USE
BP.EXPD
set to
NONE
LCD
Sound
Buzzer sounds double pulsing
tone
Buzzer sounds double pulsing
tone
None
Action
• Option A, Perform bump test:
Press and release POWER/
MODE to enter User Mode
and perform a bump test. The
instrument takes you straight
to the bump test start screen in
User Mode’s BUMP item. See
pg.37 for bump test
instructions.
NOTE
: If
Password Protection
is set ton
On
(factory setting is
Off
) using the GW-3 Config Pro
-
gram, you will be prompted for a
password to enter User Mode.
If the bump test is successful,
the screen above will not
appear again until the unit is
due for bump testing. If the
bump test is not successful,
the screen above will again
appear in the startup sequence.
• Option B, Bypass message: To
continue without performing a
bump test, press and release
AIR.
The GasWatch 3 cannot be used
until a successful bump test is
performed. Press and release
POWER/MODE to enter User
Mode and perform a bump test.
The instrument takes you
straight to the bump test start
screen in User Mode’s BUMP
item. If you don’t press
POWER/MODE, the instrument
automatically goes to the bump
test start screen after 6 seconds.
See pg.37 for bump test instruc
-
tions.
NOTE
: If
Password Protection
is set ton
On
(factory setting is
Off
) using the GW-3 Config Pro
-
gram, you will be prompted for a
password to enter User Mode.
If the bump test is successful, the
screen above will not appear
again until the unit is due for
bump testing. If the bump test is
not successful, the screen above
will again appear in the startup
sequence.
• Option A, Perform bump test:
If you want to enter User
Mode and perform a bump
test, press and release
POWER/MODE. The
instrument takes you straight
to the bump test start screen in
User Mode’s BUMP item.
• Option B, Bypass message:
To continue without
performing a bump test, wait
a few seconds for the
instrument to continue with its
startup sequence.
BP - LMT
CAL
BP - LMT
FAIL
NEXT .BP
d
0
10:42
2 . 13
2019