Beacon 110 Rig Monitor Operator’s Manual
2
Specifications
Table 1 lists specifications for the Beacon 110 Rig Monitor.
WARNING:
When using the Beacon 110 Rig Monitor, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the
Beacon 110 Rig Monitor and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the Beacon 110 Rig Monitor as described in
this manual.
Table 1: Beacon 110 Rig Monitor Specifications
Input Power
10.8 - 14.5 VDC, 1A
Construction (housing)
Fiberglass/polyester with lexan window (NEMA 4X)
Dimensions
19 in. H x 7.2 in. W x 5.6 in. D
(483 mm H x 183 mm W x 142 mm D)
Weight
13 lbs.
Environmental Conditions
•
For indoor or outdoor locations (Type 4X)
•
-20°C to 50°C (-4°F to 122°F) ambient
•
Maximum relative humidity of 80%
Relays
•
Relay contacts rated for 10A @ 115/220V~ resistive or 10A @ 30V
resistive
•
SPDT Form C except for Alarm 1 relay contact (factory wired to alarm
strobe/horn)
User Controls
•
Three push button control switches
•
One ON/OFF toggle switch
•
One push button reset switch
Signal Output
•
4 to 20 mA, 500 ohms impedance max
Sampling Method
Diffusion
Detection Range
0 to 100 PPM (parts per million) H
2
S
Response Time
90% in 30 seconds
Accuracy
± 5% of reading or ± 2 ppm H
2
S (whichever is greater)