AirLink 7032 Gas Monitor Operator’s Manual
Chapter 7: Maintenance
•
37
4. Once in Calibration Mode, the screen appears as shown below.
5. Press
Calibration Mode Active
to return to the Advanced Configuration screen.
NOTE:
If
Calibration Mode Active
is not pressed, the unit will remain in
Calibration Mode for two hours and then it will start autoscrolling through
the Real-Time Values for active channels.
6. Press the
Home
button to return to the Home screen.
NOTE:
The AirLink 7032 automatically returns to the Home screen 15 minutes after
the last button press.
7. Close the enclosure box.
8. Clamp down the enclosure latches.
Relay Test Mode
Relay Test Mode activates each relay and can be used to determine whether or not the relays and
attached alarms are functioning properly.
1. Open the enclosure box to expose the Front Panel.
2. Tap the
AirLink
logo then press and hold the
RKI
logo for 5 seconds.