AirLink 7032 Gas Monitor Operator’s Manual
Chapter 3: Installation
•
15
9. Close the enclosure box.
10. Clamp down the enclosure latches.
Connecting an AC Power Source
NOTE:
The AirLink 7032 is configured for AC or DC operation, depending on how it is
ordered. If you are using DC power as the primary power source, go to the previous
section, “Connecting a DC Power Source”.
WARNING: Verify that the power source is unplugged or turned off before you continue with
this procedure.
1. Open the enclosure box to expose the Front Panel.
2. Unscrew the two thumb-screws on the Front Panel.
3. Open the Front Panel so that the AC (Delta) Power Supply is exposed.
4. The power supply is factory-wired to the Terminal Board.
5. For versions that came with a pre-wired AC line cord: there are three wires (black, white and
green) pre-wired from the Delta power supply terminals “L” (AC Load IN), “N” (AC Neutral
IN), and “EG” (Chassis GND or Earth GND). This set of wires will be used to plug into an
AC power outlet ONCE ALL WIRING CONFIGURATIONS ARE COMPLETE.