7 Maintenance

 

7-1. Maintenance intervals and items 

 

- 33 - 

 

 
 
 
 
 

Maintenance 

 
 
 
 
This is an important instrument for the purpose of safety. 
To maintain the performance of the detector and improve the reliability of safety, perform a regular 
maintenance. 
 

7-1. Maintenance intervals and items 

 

  Daily maintenance: Perform maintenance before beginning to work. 

  Monthly maintenance: Perform maintenance on the alarm circuit (alarm test) once a month. 

  Regular maintenance: Perform maintenance once or more for every six months to maintain the 

performance as a safety unit. 

 

Maintenance 

item 

Maintenance content 

Daily 

mainten

ance 

Monthly 

mainten

ance 

Regular 
mainten

ance 

Power Supply 
Check 

Check that the power lamp lights up. 

 

 

 

Concentration 
Display Check 

Check that the concentration display value is zero. 
When the reading is incorrect, perform the zero 
adjustment after ensuring that no other gases 
exist around it. 

 

 

 

Alarm Test 

Inspect the alarm circuit by using the alarm test 
function. 

 

 

 

Gas Adjustment 

Perform the sensitivity calibration by using the 
calibration gas. 

 

 

 

Gas Alarm 
Check 

Check the gas alarm by using the calibration gas. 

 

 

 

Ground wire 
check 

Check the connection of the grounding wire (no 
looseness or corrosion) to maintain EMC 
performance. 

 

 

 

 

<About Maintenance Services> 

  We provide services on regular maintenance including span adjustment, adjustment and maintenance. 

To make the calibration gas, dedicated tools, such as a gas cylinder of the specified concentration and gas 
sampling bag must be used. 
Our qualified service engineers have expertise and knowledge on the dedicated tools used for services, 
along with other products. To maintain the safety operation of the detector, please use our maintenance 
service. 

  The followings are typical maintenance services. For more information, please contact our overseas sales 

department or local representative. 

 

Summary of Contents for SD-1DRI

Page 1: ...PT2E 2981 Smart transmitter Gas Detector Head SD 1DRI TYPE MED EX Operating Manual PT2E 196...

Page 2: ...as detector in a place where the temperature changes drastically about 10 15 or more 3 The alarm must be set within a range where the performance of the detector can be ensured In facilities compliant...

Page 3: ...How to wire 18 4 7 How to tube 22 How to Operate 23 5 1 Preparation for start up 23 5 2 Basic operating procedures 23 5 3 How to start the detector 24 5 4 Modes 25 5 5 Maintenance Mode User 26 5 6 Ho...

Page 4: ...ses or other reasons leak with the built in gas sensor The concentrations of detected gases are displayed on the seven segment LED The detector outputs gas concentration in 4 20 mA An external pump et...

Page 5: ......

Page 6: ......

Page 7: ......

Page 8: ......

Page 9: ......

Page 10: ......

Page 11: ......

Page 12: ......

Page 13: ......

Page 14: ......

Page 15: ...prevented by installing a lightning arrester before the field devices and central processing equipment For information on how to use a lightning arrester please contact the manufacturer Grounding In...

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ......

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ......

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ......

Page 29: ......

Page 30: ...ull scale is detected the sensor may be contaminated In such a case in order to ensure the reliability of the detector carry out zero calibration and span calibration properly 6 2 Fault alarm activati...

Page 31: ......

Page 32: ......

Page 33: ...m the zero adjustment after ensuring that no other gases exist around it Alarm Test Inspect the alarm circuit by using the alarm test function Gas Adjustment Perform the sensitivity calibration by usi...

Page 34: ...ks the activation of the external alarm such as a buzzer Gas Adjustment Performs the sensitivity calibration by using the calibration gas Gas Alarm Check Checks the gas alarm by using the calibration...

Page 35: ......

Page 36: ......

Page 37: ......

Page 38: ......

Page 39: ......

Page 40: ......

Page 41: ......

Page 42: ......

Page 43: ......

Page 44: ......

Page 45: ...adjustment has not been completed Check if the correct concentration span adjustment gas has been provided and check the gas adjustment tool for a gas leak If there are mistakes perform the span adjus...

Page 46: ......

Page 47: ...3 8 years 1 After replacements having need of operation check by professional service engineers To maintain the safety operation of the detector please use our maintenance service Request RIKEN KEIKI...

Page 48: ......

Page 49: ...detector If such a symptom is observed frequently take appropriate measures to eliminate the noise Sensor abnormalities E 1 The sensor is not connected or improperly connected Check if the sensor cab...

Page 50: ...ent When the environment temperature etc changes suddenly the detector cannot adjust to it and is affected by it In some cases the detector triggers an indication alarm Because the detector cannot be...

Page 51: ...alog transmission in common with power supply power supply signal common or Two wire analog transmission digital transmission HART communication Transmission specification 4 20mADC linear load resista...

Page 52: ......

Page 53: ......

Page 54: ......

Page 55: ...ower explosive limit LEL of the combustible gas to be detected is set to 100 vol A unit used to express the percentage of a specific substance or gas in a volume of solution Calibration Adjusts the re...

Page 56: ...Revision Date issued 0 First edition 11 10 2020 1 P9 Safety information Declaration of Conformity P53 Added CH4 0 100vol to the list of gases to be detected MED specification P55 Added vol to the defi...

Page 57: ......

Reviews: