4 How to Use
4-5. How
to wire
- 17 -
<Figure of Terminal Plate>
24 VDC
DC+
1
- (Common)
2
4 - 20 mA
Sig+
3
Contact
4
Contact
5
<Specifications of Terminal Plate>
Specifications of terminal plate
Rated voltage: 250 VAC
Rated current: 12 A
However, it depends on cables to be used.
Connection conditions
Cables: 0.25 - 2.5 mm
2
Bare wire length: 8 - 9 mm
Connecting tool: Dedicated handling lever (accessory) or driver (edge 3.5 x 0.5 mm)
Grounding terminal
CAUTION
The specified bare wire length must be observed when the wire insulation is peeled off.
Improper clamping of the wire due to a shorter bare wire length may cause defective electric
conduction or heating.
Catching the wire insulation due to a shorter bare wire length may cause defective electric
conduction or heating.
Exposing the wire due to a longer bare wire length may cause defective insulation or a short circuit.
Be careful not to break up the wire. If the wire is broken up when inserted to the terminal, this may
cause defective insulation or heating.
8~9
mm
8 - 9 mm