5 How to Operate
5-5. User
mode
- 25 -
<Span Adjustment "1-1">
This is used to perform the zero adjustment.
NOTE
If the span adjustment failed since the span point was significantly fluctuated from around 20.9%, it
returns to 1-1 after FAIL rather than PASS is displayed. In this case, the span adjustment has not been
completed.
1-1
Press SET key.
Current Concentration Value Display
Press the SET key to perform the zero
adjustment.
Under Span Adjustment (CAL. is displayed)
Wait for a while until the adjustment is
completed.
Span Adjustment Completed
It automatically returns to 1-1 after PASS is
displayed.
* If the zero adjustment failed, it
automatically returns to 1-1 after FAIL is
displayed.
Return to 1-1.
Return to 1-1.