Handling the calibration gas
The carbon monoxide sensor with hydrogen compensation must be calibrated separately for carbon
monoxide and hydrogen.
If hydrogen sensitivity calibration is not performed, carbon monoxide readings may be inaccurate due to
hydrogen interference.
Due to the hydrogen compensation mechanism, carbon monoxide readings may increase temporarily if
hydrogen gas concentrations increase rapidly in the atmosphere being measured.