5. Usage Instrucions
5-4. Performing air calibration
44 / 73
CAUTION
Perform air calibration in an environment that meets all of the following conditions:
・
Pressures, temperatures, and humidity levels are similar to pressures, temperatures, and humidity levels in
the actual usage environment.
・
In fresh air
Wait for the readout to stabilize before performing air calibration.
If the temperature difference between the storage location and usage location is 15 °C or greater, turn on the
power and allow the product to adjust to ambient conditions similar to those at the usage location for about
30 minutes. After this, air calibrate in fresh air before use.
1 Hold down the AIR button in measurement
mode.
The buzzer blips once, and air calibration starts.
2 Release the AIR button once the LCD display
changes from [Air HOLD] to [AdJ RELEASE].
The display automatically returns to measurement
mode once air calibration has been successfully
completed.
O2
Summary of Contents for GW-3
Page 5: ......
Page 6: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 16: ......
Page 17: ......
Page 21: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 27: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 34: ......
Page 35: ......
Page 37: ......
Page 38: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 46: ......
Page 49: ......
Page 50: ......
Page 52: ......
Page 53: ......
Page 55: ......
Page 56: ......
Page 58: ......
Page 62: ......
Page 63: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 70: ......
Page 73: ......
Page 74: ......
Page 75: ......