6−4.

  Storage or treatment when not in use for a long period 

   
  (1) Sensor 
    Oxygen sensor is a kind of cell and will continue to consume its life even though it is not 
      used.    When detector head is not used for a long period, take out the sensor from detector 
      head and store it at following environment to lower the consumption of sensor. 
 
        ・Environment of temperature (5〜10℃) 
        ・Environment of low oxygen content   
          (Put a sensor into the container and seal off the container.)   
 
  (2) Detector head   
        Store this detector head under the following environmental condition. 
       
        ・Temperature  : 0℃〜40℃   
        ・Humidity        : 30%RH〜85%RH 
        ・Environmental condition : Environment where gas, solvent, vapor and so on are not 
                                                      existing. 

         

 
 
 
 
 
 
 
6−5.

List of recommendable spare parts for regular replacement   

 

No.  Name of parts 

Inspection frequency 

Replacement interval 

Qʼ ty/unit 

Sensor holder 

At sensor replacement

At sensor replacement 

 

 
 
 
 
 

 
 
 

 
 
 
 
 
 

             

 

・Before re-start of operation , be sure to perform gas calibration.   

Contact our authorized agent or Riken Keiki for re-adjustment including gas calibration,   

  replacement of parts, etc. 

CAUTION 

                                      *  NOTE 
The replacement interval will change depending on operating condition and it does not mean 
the guarantee period.   
Replacement interval changes according to the result of regular inspection. 

Summary of Contents for GD-F3A-1

Page 1: ...onfor accid ents andd am ag e causedb y usingthis p rod uct The ind em nificationshall b e m ad e onlyund er the warrantyp olicyofp rod ucts or p arts rep lacem ent Because this is asafety unit areg u...

Page 2: ...Alarmlevel for g as alarminstrum ent shall b e set within the p erform ance of the instrum ent Our stand ardalarmlevel is 1 8 b y volum e for the p reventionof oxyg end eficiency 3 This is not control...

Page 3: ...1 0 Ap p licab le cab le 1 0 Exam p le ofcab le connection 1 0 OPERATIONM ETHOD Prep arationb efore start up 1 2 Start up 1 2 Stop p ingthe op eration 1 2 M AINTENANCE ANDINSPECTION Insp ectionfreq ue...

Page 4: ...2 PRODUCT SPECIFICATIONS Stand ardsp ecifications 1 8 Stand ardaccessories 1 8 Op tional accessories 1 8 DETECTIONPRINCIPLE 1 9...

Page 5: ...roofconstructionfor outd oor installation Also it canb eusedas intrinsicallysafesystemincom b inationw ithzenerb arrier op tion to b e ab le toinstall inhazard ous area Id e n tif ica tio n o f e a c...

Page 6: ...e Performg ascalib rationat p rop erinterval Donot chang eorrem od el thecircuit structureofthisunit Donot d isassem b lethesensorsincethereisaelectrolyteinsid e Ifyourfing ertouchestheelectrolyte w...

Page 7: ...5 PR O D U C T FU N C TIO N Ex te r n a l d r a w in g P ro tec ti v e g rou n d C ab l e i n l et fo r U b o l t S en so r...

Page 8: ...yedterm inal for sensor leadandexternal cab le Cab le g land Entrance toleadthe cab le 8 1 1 m m fromoutsid e Hing e Hing e for front d oor Lock Lockfor front d oor M ountinghole M ountinghole tofixth...

Page 9: ...visionandg roup Perform anced ivision ia Group C 3 Relationsb etw eenallow ab leind uctance Lo andallow ab lecap acitance Co ofintrinsic safetycircuit andexternal w iringind uctance Lw andcap acitance...

Page 10: ...or head Detector head Ind icator alarm unit N on hazard ous area N on hazard ous area CVVS1 2 5 sq 2 C or eq uivalent M AX 6 0 0 m S ensor hol d er O xygen A tm osp here C ab l e gl and S hi el d ed c...

Page 11: ...atureisd rasticallychang ed Placesexceed ingop eratingtem p eratureandhum id ity M a in te n a n ce s p a ce Acertainm aintenance sp ace m ust b e securedaroundthe d etector head sothat the m aintenan...

Page 12: ...nect the op tional zener b arrier b etw eenthis unit andind icator alarmunit at non hazard ous area See b elowfig Connect the shield ingw ire tochassis g round Exam p l e of i ntri nsi cal l y safe sy...

Page 13: ...d ingw ith g round resistanceof1 0 orless U sethesp ecifiedcab le Forsysteminstallation op erationandm aintenance tak ecarenot tod am ag etheintrinsically safed esig n The p ow er of the ind icator al...

Page 14: ...tchinind icator alarmunit Be sure top erformg as calib rationat the tim e ofnewinstallation Correct read ingm ay not b e ob tainedb efore g as calib ration Sto p p in gth e o p e r a tio n Thissystems...

Page 15: ...salsonecessary to carry out g as calib ration at least every 6m onths The lawin Jap an sets forth that the unit m ust issue an alarm d uring the circuit insp ection related to alarm w hile p rovid ing...

Page 16: ...unit Stop the sup p ly of zero g as d isconnect calib ration ad ap ter and confirm that the oxyg enread ingreturns to2 0 9 vol Pe r f o r m in ga s p a n a d ju s tm e n t Pack the sp an g as fresh ai...

Page 17: ...it fromthe hold er Insert newsensor intothe hold er andp ushthe sensor until it com es tothe end Connect b lackandw hite sensor leadw ires tothe term inal strip Connect w hite w ire top ositive term i...

Page 18: ...rature 0 4 0 Hum id ity 3 0 RH 8 5 RH Environm ental cond ition Environm ent w here g as solvent vap or andsoonare not existing List o f r e co m m e n d a b le sp a r e p a r ts f o r r e g u la r r...

Page 19: ...othe oxyg en onten p erm eatedthroug hthe m em b rane Refer to 1 0 DETECTIONPRINCIPLE for d etails Safety retainer for exp losion p rotectionb y lim itingthe energ y ofelectric current w ithin anon ig...

Page 20: ...le cab le O D 8 1 1 m m Op eratingtem p erature 1 0 4 0 Op eratinghum id ity Below9 0 RH Non cond ensing Structure W all m ounting 2 B 5 0 A p ole m ounting Kindofexp losionp roof Intrinsically safe w...

Page 21: ...teat aratep rop ortional tothep artial p ressureofoxyg en Theoxyg enreactsinthe cell and p rod uces a current p rop ortional to the oxyg en content The current d evelop s a voltag e across a tem p era...

Reviews: