RIGOL
RP1003C/1004C/1005C/1006C User Guide
7
waveform measurement instrument. Turn the collar until it clicks, and check
that it is locked securely.
CAUTION
When disconnecting the output connector from the waveform
measurement instrument, be sure to release the lock before pulling out
the connector. Forcibly pulling out the connector without releasing the lock,
or pulling the cable, can damage the terminator.
If using BNC-banana plug adapters or similar to connect to input terminals
other than BNC connectors, make sure the polarity is correct.
Do not demagnetize while the current probe is clamping a conductor to be
measured. Demagnetizing causes current to flow into the conductor,
which may damage parts in the circuit to be measured.
RP1003C/RP1004C
RP1005C/RP1006C
Check that the conductor being measured is not clamped when supplying
power to the current probe for the same reason. Demagnetized
waveforms are generated when supplying electric power.
4.
Make sure the current sensor is locked (for RP1003C and RP1004C, the slide
switch should be in the LOCK position; for RP1005C/RP1006C, LOCK should be
displayed on the slide switch and UNLOCK should disappear).
5.
Press the demagnetizing switch (DEMAG) on the terminator.
6.
Turn the zero adjustment dial on the terminator to adjust the trace to the zero
position.
Collar
×
×
Demagnetizing
Demagnetizing
Summary of Contents for RP1003C
Page 1: ...User Guide RP1003C 1004C 1005C 1006C Current Probe Sept 2020 RIGOL TECHNOLOGIES CO LTD ...
Page 2: ......
Page 10: ......
Page 29: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 19 RP1005C RP1006C ...
Page 31: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 21 RP1005C RP1006C ...
Page 33: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 23 RP1005C RP1006C ...