RP1003C/1004C/1005C/1006C User Guide
To Use the Current Probe
Before using the current probe, make sure to refer to
Safety Precautions
.
Preparations for Measurement
1.
Prepare one RP1000P power supply and one waveform measurement
instrument for waveform measurement.
CAUTION
Before turning the power supply on, make sure that the voltage matches that
indicated on the rear panel of the RP1000P. Connection to an improper supply
voltage may damage the RP1000P and present an electrical hazard.
2.
Turn the power switch off and connect the power cord.
3.
Connect the power plug of the current probe to the power receptacle of the
RP1000P.
4.
Turn the RP1000P power switch on, and check that the front panel power
indicator lights on.
5.
Wait at least 30 minutes after turning on the current probe. Immediately after
power is supplied, offset drift may increase due to the effects of self-heating of
the device and other factors. To ensure accurate measurement, wait at least 30
minutes after turning on the current probe before performing the
measurement.
Demagnetizing and Zero Adjustment
1.
With the waveform measurement instrument input at ground, adjust the trace
to the zero position.
2.
Set the input coupling of the waveform measurement instrument to DC.
3.
Connect the output connector of the current probe to the input connector of the
Summary of Contents for RP1003C
Page 1: ...User Guide RP1003C 1004C 1005C 1006C Current Probe Sept 2020 RIGOL TECHNOLOGIES CO LTD ...
Page 2: ......
Page 10: ......
Page 29: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 19 RP1005C RP1006C ...
Page 31: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 21 RP1005C RP1006C ...
Page 33: ...RIGOL RP1003C 1004C 1005C 1006C User Guide 23 RP1005C RP1006C ...