Trigger Parameter Setting
You can refer to
to set the trigger holdoff and
noise rejection under this trigger type.
8.7.10
Delay Trigger
In Delay trigger, you need to set Source A and Source B. The oscilloscope triggers
when the time difference (∆T) between the specified edges (Edge A and Edge B) of
Source A and Source B meets the preset time limit, as shown in the figure below.
Edge A and Edge B must be two neighboring edges.
T
Source A
Source B
Edge A
Edge B
Figure 8.18 Delay Trigger
Trigger Type
Click or tap the drop-down button of Type to select "Delay".
Figure 8.19 Delay Trigger Setting Menu
After a trigger type is selected, the current trigger setting information (including
trigger type, trigger source, and trigger level) is displayed in the trigger information
Triggering the Oscilloscope
HDO1000 User Guide