8.7.8
Runt Trigger
The runt trigger sets the oscilloscope to trigger pulses that cross one trigger level but
not another, as shown in the figure below.
Lower Limit of the
Trigger Level
Upper Limit of the
Trigger Level
Negative Runt Pulse
Positive Runt Pulse
Figure 8.15 Runt Trigger
Trigger Type
Click or tap the drop-down button of Type to select "Runt".
Figure 8.16 Runt Trigger Setting Menu
After a trigger type is selected, the current trigger setting information (including
trigger type, trigger source, and trigger level) is displayed in the trigger information
label at the top of the screen, as shown in the figure below. The information will
change based on the trigger settings.
Triggering the Oscilloscope
HDO1000 User Guide