Limitations
Specification
- RIGHTEST Blood Glucose Monitoring System is not intended for serum or plasma testing.
- Inaccurate test results may be obtained at altitudes greater than 3,048 meters(10,000 feet) above
sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- The glucose test may be invalid in the presence of abnormally high concentration:
Ascorbic acid
≧
5 mg/dL (0.28 mmo/L)
Uric acid
≧
20 mg/dL (1.19 mmol/L)
Xylose
≧
20 mg/dL (1.33 mmol/L).
- Not for screening or diagnosis of diabetes mellitus.
- Regarding potential electromagnetic or other interference, keep meter out of electromagnetic
radiation sources.
Measurement Technology
Measuring Range
Test Time
Memory Capacity
Power Saving
Operating Temperature
Operating Relative Humidity
Power Supply
Battery Life
Dehydrogenase Electrochemical Sensor
10 - 600 mg/dL (0.6 - 33.3 mmol/L)
5 seconds
1,000 blood glucose test results with date and time
6 - 44°C (43 - 111°F)
10 - 90%
2 CR2032 Batteries (3 Voltage)
Approximately 1,000 standard tests
Turns off automatically 2 minutes after last user action. To
turn off manually, press the button " " for 4 seconds.