10
Technical Specifications
Oxygen Saturation Simulation (PULS-R)
Range
30 to 100%
Repeatability
± 5%** of reading between 30-59% SPO2
± 3% of reading between 60-99% SPO2
± 3% of reading between 90-100% SPO2
Accuracy of simulation when used with the corresponding R-curves
*Based on using the same probe and monitor setup
**Note that some monitor types might not be able to display low range sats
Heart rate
30-300 bpm***
Accuracy
± 1 bpm
Compatibility
Beijing Choice, Criticare, GE Tuffsat,
Masimo,
Mindray,
Nellcor, Nellcor Oximax, Nihon Kohden,
Nonin,
Novametrix, Philips / HP
***Subject to monitor capability
Operation
Battery cell, in-situ charge
Mains supply
230V/110V AC +/- 10% 50/60Hz
Battery life
Approx. 11 hours testing
Weight
<1 kilo
Dimensions
26 x 10 x 5.5 cm
Operating conditions
10-30'C, 0-90% RH - NC
Storage environment
-15' - +60'C
Environmental Protection
IP 40