45 |
Page Rigel Medical
SP-SIM
user Manual V2.2
6. Use the Up & Down Keys to navigate the menu and the Left & Right Keys to
open the drop down boxes [<>]. Select the required input and confirm with the
button (F4). Each variable entered during the automatic test will
automatically be added to the list for future use. Trace Variables that are de-
activated (set to NO in the Setup menu, section 4.2) will not appear during the
automatic tests.
Note that a maximum of 40 individual entries are available for each Trace
Variable. Additional entries will overwrite on a first in first out basis.
When all information is provided, press the
button (F4) to start the safety test.
A default visual inspection is provided at the start of each sequence.
In the following example, a simulation is done in semi automatic mode where the user
can enter the results in between each test; An initial simulation is done at 98% foloowed
by a chronometer test and capturing of alarm (using a custom test, text input)