TheraTouch® CX2/CX4/EX4
25
APPLICATION INFORMATION
NAVIGATION
Depending on the electrode size, use two
or three wraps to maximize the contact
surface.
After use, clean per rubber electrode
manufacturer instructions.
Self-Adhesive Electrodes
Self-adhesive electrodes have higher series
impedance than flexible rubber electrodes.
This can cause the stimulator to terminate
treatment at higher current amplitudes.
When this occurs it is recommended to
continue the treatment with flexible rubber
electrodes, properly moistened with a
conductive medium.
Electrodes smaller than 3" x 5" are not
recommended for use with currents that
contain a DC component.
Connection and Disconnection Reactions
Constant Current (CC) output
characteristics may cause unpleasant
connection and disconnection reactions if
the electrodes are not securely placed or
lose contact with the skin. Make sure the
current amplitude is set to 0 mA when you
apply or remove the electrodes.
This manual contains instructions for
the TheraTouch® CX2, CX4 & EX4. For
TheraTouch EX4 set-up instructions, refer
to the Navigation — Electrotherapy section
of this manual.
For TheraTouch CX2/CX4 set-up
instructions, refer to the Navigation
Section of this manual, which is divided
into three main set-ups:
•
Electrotherapy only
•
Ultrasound Therapy
•
Combination Therapy
ELECTROTHERAPY SETUP – CLINICAL PROTOCOLS
Home
The Home Menu gives access
to all functions of the unit.
•
Select the Electrotherapy button.
Electrotherapy
The Electrotherapy Menu gives you
access to the following functions:
• Clinical Protocols
• Favorites
• Manual Operation
•
Select Clinical Protocols button.