TheraTouch® CX2/CX4/EX4
24
ULTRASOUND TREATMENT SCREEN
1.
Remaining Treatment Time
2.
Output Value
3.
Unit of Output Value (W/cm
2
)
COMBINATION THERAPY SCREEN
1.
Ultrasound Output Indicator
2.
Electrode Output Indicator
3.
Remaining Treatment Time
Electrotherapy
4.
Output Value
5.
Unit of Output Value (mA)
Ultrasound
6.
Output Value
7.
Unit of Output Value (W/cm
2
)
2
1
3
7
6
3
1
2
4
5
DEVICE OPERATION
APPLICATION INFORMATION
ELECTROTHERAPY
CAUTION
•
Connection of accessories other than
the ones specifi ed by the manufacturer
can adversely aff ect the safety of the
patient and correct functioning of
the equipment, and is therefore not
permitted.
•
To prevent infection, electrodes should
not be used on broken skin.
Before Treatment
•
Check the patient for Contraindications
as noted under Intended Use section of
this manual.
•
Test the heat sensitivity of the treatment
area.
•
Clean the treatment area. Any areas
with excess hair should be clipped or
shaved to ensure good conductivity.
Rubber Electrodes (Flexible)
If your clinic is using fl exible rubber
electrodes, a conductive medium should
be used, such as conductive gel, etc.
Prior to initial use, put enough conductive
medium to saturate the rubber electrode
or to cover the entire surface area of the
rubber electrode. Only use a conductive
medium specifi cally approved for
electrotherapy stimulation.
DO NOT
use a
medium with an alcohol-based content.
Place the electrodes on the patient and
apply an elastic wrap around the electrode
to secure it in place.